How to record instrument calibration

By pressing play, YouTube (Google) loads the video and may store cookies on your device. Privacy notice

The Calibration page lists every device a critical limit is judged with: probes, fixed thermometers, pH meters, scales and timers. For each device you record when it was checked against a reference, and what was done when it read wrong.

Before you start

A calibration compares the device's reading with a reference taken at the same moment. The difference is the deviation. The tolerance is how far the deviation may go before the device has failed. When a device fails, every reading it gave since its previous calibration is in doubt, and so is the product judged with those readings.

  • You are an owner, admin or member. Contributors and viewers see this page read only.
  • Your package includes ISO 22000.
  • It helps to set the critical control points first, so that you can link each device to the points it judges.
  • Have a reference ready: iced water, a certified thermometer or a test weight.

Step by step

1. Open the page

Open Calibration under Food Safety in the menu. The list shows the devices of every site, each with its site.

2. Add a device

Press Add a device. Once the list has devices, the button reads Add device.

The device is created at once as New device, and its row opens.

3. Describe the device

Fill in the fields. Each one is saved as you type.

a. Name, and Serial / asset tag: what is written on the device itself.

b. Kind: Temperature probe, Fixed thermometer, pH meter, Scale, Timer or Other.

c. Location.

d. Calibrate how often, for example Monthly.

e. Tolerance, with its unit. Leave it empty if it is not decided yet.

Under Used to judge which control points, press the chip of each control point the device measures. A failed calibration then lists those points.

5. Record a calibration

a. Press Record calibration on the device.

b. Enter the Device reading and the Reference, both read at the same moment.

c. Check the Result, and change it if you judged otherwise. With a tolerance set, it is proposed from the deviation.

d. Fill in Calibrated by.

e. Press Record.

6. Deal with a failed device

a. If the result is Out of tolerance, fill in What happens to the affected product. The calibration is not recorded without it.

b. Press Record. A corrective action is raised. The card Readings in doubt appears at the top of the page.

c. Press Open corrective actions to follow it up.

The device shows "out of use until a new calibration passes". The next passing calibration clears this.

7. Correct a mistake

Calibration history is append-only. To correct a calibration entered wrong, record a new one.

8. Retire or delete a device (optional)

Retire device takes the device off the list at once, with no confirmation. This page has no way to bring it back.

Owners and admins also see Delete, which asks for confirmation first.

Example from a kitchen

What Krateon saves

The device with all its fields, for the site where it was added. Each calibration is added to the device's history: the date, the reading, the reference, the deviation, the tolerance, the result, who calibrated it and, on a failure, the answer about the product.

A failed calibration also saves a corrective action, numbered in the form CAR-001 and with a target date, and a cause for review.

Where it shows up later:

  • HACCP verification, row "Measuring instruments and thermometers".
  • HACCP dossier, section 8 (Calibration of instruments: programme and records).
  • Calendar. "Calibration due: {x}" when a calibration falls due.
  • Corrective actions and System origin and upkeep, after a failure.

If something goes wrong

The quoted texts in the first column appear in the app exactly as shown.

What you seeWhat it meansWhat to do
"This page is not enabled."ISO 22000 is not switched on for your organisation.Contact support at hello@krateon.com.
"Included in {what}" with the button Ask for itYour package does not include ISO 22000. The page is read only.Press Ask for it.
"The records could not be loaded."The page could not read the register.Press Try again.
An empty card with the button Add a deviceNo device has been added yet.Press Add a device and do step 3.
"Required: {fields}"A required field of the calibration is empty.Fill in the fields it names and press Record again.
"Enter a number"The reading or the reference is not a number.Type a number, without the unit.
"This device has no tolerance recorded, so nothing is proposed. Decide the result yourself, or set a tolerance first."The device has no tolerance.Choose the result yourself, or fill in Tolerance first.
"No control points exist yet."No control point exists to link to.Set the control points first, then link the device.
Calibration overdueThe next calibration date has passed.Calibrate the device and do step 5.
Never calibratedThe device has no calibration yet.Do step 5.

Sources:

  • No EU or Greek law sets a kitchen probe tolerance; Regulation (EC) No 37/2005, Article 2 (quick-frozen food recorders only): https://eur-lex.europa.eu/eli/reg/2005/37/oj/eng